GT300-AQ/VOC provides CIAQ (Composite Index of Air Quality) suitable to use in various applications and detect air quality and various odors such as swage-sludge gas, methane, smoke, smelling breath, particular gas, VOCs, THC and so on. GT300-AQ/VOC make the detecting easily and fun.
GT300-AQ/VOC can detect hydrocarbon compounds in the air, especially high sensitive of hydrogen and transfer the CIAQ (Composite Index of Air Quality) shown on the display. CIAQ define that fresh air and/or zero air is "1". According to concentration of hydrocarbon compounds and hydrogen in the air, the CIAQ is unlimited in theory. Applicant in the living, when concentration of hydrocarbon compounds over some limit in the air, the sense of smell will not have multiple effects, therefore CIAQ is designed in reasonable range in living. Please refer table as below:
Environment and Application | CIAQ of GT300-AQ/VOC |
Fresh air and/or Zero air | 1 |
Living room | 1 - 8 |
Super market, Convenient store | 10 - 20 |
Office | 5 - 20 |
Road side or Bus stop of downtown | 10 - 30 |
Public lavatory | 10 - 40 |
Packing lot or Tunnel of automobile | 10 - 40 |
Hospital or Sickroom | 15 - 25 |
Smoking room of airport | 30 - 90 |
Temple | 30 - 50 |
Fart (close detecting) | 50 - 100 |
Smelling breath (Around blow 5 sec continuously) |
20 - 35 |
Opening sewage | 20 - 40 |
In above table except zero air CIAQ is absolutely "1", the other index may be different with user actual detecting, but the higher CIAQ means the more pollutant or hydrocarbon compounds exist. Basically GT300-AQ is more sensitive than GT300-VOC in detecting hydrogen and smoke smell. But it cannot expose in the environment of high concentration VOC for a long time. If detecting targets are VOC and LPG etc, the GT300-VOC is more suitable.
Part of GT300-AQ / VOC functions
Odor detecting
The smell and sense of human is difficult to define, some of scent sense good but some persons might not accept the smell. Although everybody has various different sense of smell that cannot define. Anyway, the smell depends on the concentration of hydrocarbon compounds in the air.
Smoke detecting
GT300-AQ is very sensitive of hydrogen which generated from smoking or incomplete combustion. Therefore, GT300-AQ is suitable for air-conditioner management and non-smoking surveillance. Besides, GT300-AQ fire detective performance is superior to normal smoking detector.
Leakage detection
of
LPG and LNG
GT300-AQ/VOC is very sensitive and suitable to detect leakage of Natural Gas (Methane / City Gas) and LPG (Propone).
Detection of
VOCs emission
It is hard work to find out the source of emission of VOCs? GT300-AQ/VOC will help user to detect the source of leakage easily.
Environmental index
CIAQ is objective index of environmental either living in most application. Due to smell and sense of human is difficult to define but
GT300-AQ/VOC is non-subjective instrument that makes comparison and evaluation of environmental.
Kiss Index
In order to win some person favor, pay attention to the politeness is necessary. Bad breath affects most people at some stage of their life, but
GT300-AQ/VOC is simple detector to prevent it and help you win the favor.
Air-conditioner management and air cleaner
There are various factors such as smoking, odor of rotten food, using volatile solvent, using liquid eraser, bad ventilating and so on, cause polluted air and high CIAQ. In order to improve air quality, we may use air cleaner. But nobody knows what is the air cleaner performance?
GT300-AQ/VOC will help user to know the air quality and air cleaner is effect or not.
VOCs,
THC and Particular gases detecting
It is not easy to buy suitable gas detector for some of particular gases. In the case, GT300-AQ/VOC might be a substitute and solution. There are some notices that the concentration of target gas should be 10~1000 ppm, balance gas should be air without interfered gases or exist few stable interfered gases.
For example GT300-AQ/VOC is more sensitive and practical than other professional detector to detect exist less than 100 ppm of hydrogen. If the gas cannot measure by diffusion or high concentration that out of CIAQ range, dilute detection is possible, put
GT300-AQ/VOC in the bottle and takes the sample gases then inject bottle and read the CIAQ. Neither diffusion or dilute, Silicon compounds gases is not suitable to use
GT300-AQ/VOC, e.g. SiH4.
The best partner
of canary |
|
More application
GT300-AQ/VOC detector has various applications, i.e. CIAQ reaction of fart (near measure) can find out who fart. Besides, above introduction others application may be develop by user.
A fast method and cost-saving of measuring THC and VOCs
It takes time and cost to measure THC (Total Hydrocarbons), VOCs (volatile organic compounds) or some of particular gases when use FID or GC. Even time is very critical but nothing can change except waiting while use them. Now, GT300-VOC changed this situation, it makes fast and cost saving of measuring.
GT300-VOC can detect hydrocarbon compounds and VOCs in the air, and transfer the CIAQ (Composite Index of Air Quality) shown on the display. CIAQ define the fresh air and/or zero air is "1". According to concentration of hydrocarbon compounds and hydrogen in the air, the high CIAQ means high concentration of hydrocarbon compounds. This characteristic can transfer the concentration of gas in ppm or mg/cubic meter.
CIAQ is created by Gastech that is very strange for every one with perplexed. But its convenient and economic features catch users eyes and make some inquiries. For example, what is change rate from CIAQ to hydrogen or THC? Actually, interfered gases and different targets influence the value of CIAQ. Therefore, no absolute change rate can be used. There is same story in wastewater control, use COD instead of BOD5 for some purpose and fast. But the change rate should be found on past data and comparison. Due to different case of wastewater content is different that the change rate of case A is not suitable in case B normally.
It is an actual case study that applies FID and GT300-VOC together and made a table as below.
THC (Total Hydrocarbons), mg/cubic meter | CIAQ (Composite Index of Air Quality) |
14.9 | 15 |
29.9 | 22 |
59.9 | 32 |
119.8 | 43 |
149.8 | 47 |
199.7 | 54 |
249.6 | 58 |
299.6 | 60 |
399.4 | 67 |
499.3 | 71 |
According the pair of data marks them and makes calibration curve on logarithmic graph paper, which vertical axis is CIAQ and horizontal axis is THC. The curve is nearby a line that means no more data will get high accuracy while makes calibration curve.
It will be convenient to measure THC after users have the calibration curve. For example, GT300-VOC read out 40 CIAQ and lookup 100 mg/cubic meter THC nearly in same environment. In the past, to know the environmental THC in urgent case, overslept the timing always after prepare the FID or sample tools. To use GT300-VOC will be no more regret in the case, user can measure THC and get result in anytime immediately.
Same method, GT300-VOC can be used as hydrogen, particular gases and VOCs detector. The unit of horizontal axis can be mg/cubic meter or ppm. It's a cost-saving and fast method of measuring.
Specification
Features \ Model |
GT300-AQ |
GT300-VOC |
Sensor |
Semiconductor |
Semiconductor |
Major Detection | Hydrogen, Smoke | VOCs, THC |
Available / Target gases |
Air Quality VOCs emission Smoke Odor Hydrogen Ethanol Iso-butane Carbon monoxide Methane / LNG Hydrocarbon Compounds |
Air Quality VOCs emission Smoke Odor Toluene Hydrogen sulfide Ethanol Ammonia Hydrogen Hydrocarbon Compounds |
Major Application | Air Quality | VOCs, THC |
Sampling frequency | Continuance | Continuance |
Alarm |
High alarm (Resetable) |
High alarm (Resetable) |
Audible alarm / Alarm message |
Available |
Available |
Backlight alphanumeric LCD |
Available |
Available |
Standard language |
English |
English |
System clock |
Year, month, day, hour, minute |
Year, month, day, hour, minute |
Battery management |
Available |
 Available |
Dimensions |
131mm x 46mm x 25.5mm |
131mm x 46mm x 25.5mm |
Weight |
130g |
130g |
Battery |
Built-in 2000mAH rechargeable lithium-ion battery |
Built-in 2000mAH rechargeable lithium-ion battery |
Standard accessory |
Adaptor |
Adaptor |
International standard met |
CE |
CE |
Package size |
208 mm x 156 mm x 68 mm |
208 mm x 156 mm x 68 mm |
Package outline |
(Specification maybe changed without notice)
|
Download Catalog of GT300-AQ/VOC 334 KB and 7 Pages, Feb. 10, 2003 |
GASTECH is a professional manufacturer in various gas detectors, carbon monoxide detector, carbon dioxide detector, methane detector, oxygen detector, odor detector, VOC detector, hydrogen sulfide detector and various gas analyzer, carbon monoxide analyzer, carbon dioxide analyzer, methane analyzer, oxygen analyzer, odor analyzer, VOC analyzer, hydrogen sulfide analyzer also. The combustible gases detector of model GTF200-FL is suitable to detect Methane, Propane, Butane, Iso-Butane, n-Pentane, n-Hexane, n-Heptane, Iso-Octane, Methanol, Ethanol, Iso-Propanol, Acetone, Toluene, Ethyl Acetate, Hydrogen, Cyclohexane, Acetylene, Ethylene etc. It could be as a Methane Detector, Propane Detector, Butane Detector, Iso-Butane Detector, n-Pentane Detector, n-Hexane Detector, n-Heptane Detector, Iso-Octane Detector, Methanol Detector, Ethanol Detector, Iso-Propanol Detector, Acetone Detector, Toluene Detector, Ethyl Acetate Detector, Hydrogen Detector, Cyclohexane Detector, Acetylene Detector, Ethylene Detector etc. GT200-3G, GT200-2G, GT220, GT300-AQ, GT300-VOC, GTF200-OX, GTF200-CO, GTF200-FL, GTF220, GTF300-AQ, GTF300-VOC and Portadetector are our product and brand name. Welcome tour our website www.gastech.com.tw